Milwaukee Mi180 Multi Parameter Meter

home  -   contact  -   site map
 

Milwaukee Mi180 Multi Parameter Meter

Milwaukee Mi180 pH/ORP/EC/TDS/NaCl/Temperature Laboratory Bench Meter

The Milwaukee Mi180 measures 6 different parameters: pH, ORP, EC, TDS (Total Dissolved Solids), % of NaCl and Temperature in a variety of ranges. pH calibration can be performed in 3 points selectable between 7 memorised buffers, to provide a very accurate calibration curve even when testing different samples, where very wide differences in pH can be found.

The auto-ranging feature for EC and TDS measurements automatically sets the resolution suitable to the tested sample. All measurements can be temperature compensated at 20°C or 25°C and the compensation coefficient is selectable by the user.

The automatic temperature compensation can also be disabled for measuring the actual conductivity value. The stability indicator on the LCD ensures accuracy. Conductivity readings are performed with the 4-ring probe supplied with the meter. The GLP feature allows users to store and recall data on system status. The Mi180 is PC compatible through an RS232 port or USB.

Ordering information:

Mi180 is supplied complete with:

  • MA917B/1 pH electrode
  • MA813D/1 EC/TDS/NaCl/Temperature probe
  • MA831R Temperature probe
  • MA9315 Electrode holder
  • M10004 pH 4.01 sachet buffer solution
  • M10007 pH 7.01 sachet buffer solution
  • M10010 pH 10.01 sachet buffer solution
  • M10016 sacher electrode cleaning solution
  • Mi5200 Application software
  • MA9350 RS232 connection cable with 2 metre cable
  • Graduated pipette, 12 VDC adaptor and instruction manual

Specifications:

Range        pH

                 mV

                 EC

 

                

                 TDS

 

                

                 NaCl

                 Temp.

-2.00 to 16.00 pH; -2.000 to 16.000 pH

± 699.9 mV: ± 2000 mV

0.0       to 29.99 μS / cm; 30.0 to 299.9 μS / cm; 300 to 2999 μS / cm

3.0 to 29.99 mS / CM; 30.0 to 200.0 mS / CM; up to 500.0 mS / cm (uncompensated EC*)

0.0 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt); up to 400.0 g/L actual TDS (with 0.80 factor)

0.0 to 400.0%

-20.0 to 120.0°C / -4.0 to 248.0°F

Resolution pH

                   mV

                   EC

                  TDS

                   NaCl

                   Temp.

0.01 pH; 0.001 pH

0.1 mV; 1 mV

0.01 μS / cm; 0.1 μS / cm; 1 μS / cm; 0.01 mS / cm; 0.1 mS / cm

0.01 mg/L; 0.1 mg/L; 1.0 mg/L; 0.01 g/L; 0.1 g/L

0.1%

± 0.1°C / 0.1°F

Accuracy     pH

                    mV

                    EC

                   TDS

                    NaCl

                   Temp.

 

± 0.01 pH; ± 0.002 pH

± 0.2 mV; ± 1 mV

± 1% of reading ± (0.05 μS / cm or 1 digit)

± 1% of reading ± (0.03 ppm or 1 digit)

± 1% of reading

±0.4°C / ± 0.8°f

Rel. mV offset

± 2000 mV

Calibration pH

 

                    EC

 

                    

                    NaCl

                    Temp.

1, 2 or 3 points calibration, with 7 memorised buffers (pH 1.68, 4.01, 6.66, 7.01, 9.18, 10.01 and 12.45)

1 point slope calibration with 6 memorised solutions (84 μS / cm, 1413 μS / cm, 5.00 mS / cm, 12.88 μS / cm, 80.0 μS / cm and 111.8 mS / cm

1 point, with MA9066 solution

2 point, at 0 and 50°C / 32 and 122°F

Temperature compensation

automatic or manual, from -20.0 to 120.0°C / -4.0 to 248.0°F

Temperature coefficient

selectable from 0.00 to 6.00%/°C (EC and TDS only)

pH electrodes and temp. probe

MA917B/1 & MA831R (included)

EC/TDS/NaCl/Temp. probe

MA813D/1 (included)

TDS factor

0.40 to 0.80 (default value is 0.50)

Log on demand

up to 50 samples on each range (pH, mV, EC, TDS, NaCl)

GLP

last pH, ec, NaCl calibration data

PC interface

RS232 / USB Opto-isolated

Environment

0 to 50°C / 32 to 122°F; max. RH 95%

Input Impedence

1012 Ohm

Power supply

12 VDC power adaptor (included)

Dimensions

230 x 160 x 95 mm.

Weight

0.9 kg.

(*) Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation

Accessories

MA9001

pH 1.68 buffer, 230ml. bottle

MA9004

pH 4.01 buffer, 230ml. bottle

MA9006

pH 6.86 buffer, 230ml. bottle

MA9007

pH 7.01 buffer, 230ml. bottle

MA9009

pH 9.18 buffer, 230ml. bottle

MA9010

pH 10.01 buffer, 230ml. bottle

MA9112

pH 12.45 buffer, 230ml. bottle

MA9012

Refilling solution for pH electrode, 230ml. bottle

MA9015

Electrode storage solution, 230ml. bottle

MA9016

Electrode cleaning solution, 230ml. bottle

MA9112

pH 12.45 buffer, 230ml. bottle

MA9060

12880 μS / cm calibration solution, 230ml. bottle

MA9069

5000 μS / cm calibration solution, 230ml. bottle

MA9061

1413 μS / cm calibration solution, 230ml. bottle

MA9063

84 μS / cm calibration solution, 230ml. bottle

MA9065

111.8 mS / cm calibration solution, 230ml. bottle

MA9066

100% NaCl calibration solution, 230ml. bottle

MA9310

12 VDC adaptor, 220 V

MA9311

12 VDC adaptor, 110 V

MA9315

Electrode holder

MA917B/1

Double junction refillable pH electrode

MA813D/1

EC/TDS/NaCl/Temperature probe with DIN connector and 1 metre cable

MA921B/1

Double junction, gel filled ORP electrode

MA831R

Temperature probe

MA9350

RS232 connection cable with 2 metre cable


 
Search
Search Site:
Make an Enquiry
Brochures

   Progen Focus 2017

 

To view the brochures you will need to have Acrobat Reader
 
Site Build and Hosting by Wizzy Design Ltd.  |   Design By Website Optimisation  |   Powered By Apex CMS ©