Cruma Ductless Fume Hoods

home  -   contact  -   site map
 

Cruma Ductless Fume Hoods

Progen Scientific can supply the CRUMA portable fume hoods designed in compliance with the standards: BS 7258, EN ISO 9001:2000, AFNOR NF X 15-211 (Class II), CSA Z 316.5, EN-1822, EN-779 and EN-141.

Features of CRUMA ductless fume hoods for gases:

  • Illumination
  • Liquid retention tray (2 litre capacity)
  • Safety glass work surface
  • Anemometer for continuous control of the air face velocity and inside the fume hood
  • Alarm in case of fan failure
  • Door open alarm
  • System to detect the saturation of the filter using colorometric reactive tubes)
  • Incorporation of a digital timer with audible and visual indications, warning every 60 hours of the need to check the saturation level of the filter
  • Structure made from 2mm. steel plate with acid resistant polymerised Epoxy coating
  • Side panels and front doors of transparent 8mm. thick polymethyl methacrylate, highly resistant to fire and acids
  • Low noise level centrifugal fans (50 dB), placed after the filter to prevent corrosion
  • Fully programmable electronic circuit that can be customised on request
  • 3 year guarantee

CRUMA Ductless Fume Hoods range

Each size is available in two models:

  • G for gases and vapours
  • GS for gases, vapours, solids and aerosols

Common specifications of all models:

  • Nominal air flow (cu.metre/hour): 205
  • Average face velocity (metre/second): 0.50
  • Voltage - frequency: 220V - 50Hz
  • Noise level: 48 dB

CRUMA 1010

G

GS

Internal hood volume (cu.m.)

0.458

0.458

Air renewals per minute

6.4

4.4

Total elect. power consumption

105W

105W

Illumination: Philips PLL kit

55W

55W

Internal dimensions (WxDxHmm.)

975 x 560 x 940

975 x 560 x 940

External dimensions (WxDxHmm.)

1000 x 600 x 1225

1000 x 600 x 1225

 

CRUMA 990

G

GS

Internal hood volume (cu.m.)

0.361

0.361

Air renewals per minute

8

5.5

Total elect. Power consumption

105W

105W

Illumination: Philips PLL kit

55W

55W

Internal dimensions (WxDxHmm.)

975 x 560 x 740

975 x 560 x 740

External dimensions (WxDxHmm.)

1000 x 600 x 1030

1000 x 600 x 1030

 

CRUMA870

G

GS

Internal hood volume (cu.m.)

0.287

0.287

Air renewals per minute

10

7

Total elect. power consumption

86W

86W

Illumination: Philips PLL kit

36W

36W

Internal dimensions (WxDxHmm.)

775 x 560 x 740

775 x 560 x 740

External dimensions (WxDxHmm.)

800 x 600 x 1030

800 x 600 x 1030

 

CRUMA 670

G GS

Internal hood volume (cu.m.)

0.181

0.181

Air renewals per minute

16

12.4

Total elect. Power consumption

74W

74W

Illumination: Philips PLL kit

24W

24W

Internal dimensions (WxDxHmm.)

575 x 560 x 630

575 x 560 x 630

External dimensions (WxDxHmm.)

600 x 600 x 930

600 x 600 x 930

 

CRUMA 650

G GS

Internal hood volume (cu.m.)

0.128

0.128

Air renewals per minute

22.8

15.6

Total elect. power consumption

74W

74W

Illumination: Philips PLL kit

24W

24W

Internal dimensions (WxDxHmm.)

575 x 560 x 445

575 x 560 x 445

External dimensions (WxDxHmm.)

600 x 600 x 730

600 x 600 x 730

ECO² Fume Hood

 
Designed to meet two main criteria, Economy and Ecology (hence ECO²), for the handling of chemical reagents and compounds at room or moderate temperature in all kinds of laboratories.
 
Features:
·        External dimensions: 780W x 600D x 950H (1010H for GS) mm.
·        Internal dimensions: 770W x 560D x 760H mm.
·        Noise level: 50dB
·        Solid sides and rear wall (optional transparent for educational)
·        Anemometer
·        Filter saturation detection system
·        Volume of treated air / hour (m³/hr.): 175 (170 for GS)
·        Average face speed of air (m³/hr.): 0.50
·        Internal hood volume (m³): 0.287
·        Air removals in the hood per minute: 10.2 (9 for GS)
·        Illumination (optional) Phillips PLL kit: 36W
·        Average assembly time: 20 – 30 minutes
 
Ordering information:
 
CRECO²G / GS           CRUMA ECO² Fume Hood supplied with pre-filter                 
 
Main filters (must be ordered with fume hood)
A001               Type A filter for G hood                                                                      
AD001             Type AD filter for GS hood                                                                             
B001                Type BE filter for G hood                                                                    
BD001             Type BED filter for GS hood                                                               
F001                Type F filter for G hood                                                                       
FD001             Type FD filter for GS hood                                                                  
K001               Type K filter for G hood                                                                      
KD001                        Type KD filter for GS hood                                                                 
MX001            Type ABEK filter for G hood                                                               
MXD001         Type ABEKD filter for GS hood                                                          
MD001            Type MD filter for GS hood                                                                 
P001                Type P (HEPA) filter for GS hood                                                       
Options:
PLL                 Illumination (Phillips PLL kit)                                                    
TRAY              Liquid retention tray and safety working surface                        
KIT001            Gas detection kit                                                                                  
KIT002            Gas detector tubes (pack of 10) Specify gas tested on order
PRE4               Pre-filters (pack of 4)                                                                          

PRE10             Pre-filters (pack of 10)

Choice of Filters for the CRUMA Ductless Fume Hoods

  • Type A: ideal for ketones, ethers, alcohols, cyclic solvents. It may also be used for some inorganic acids provided they are in low proportion as the activated carbon is not impregnated and any excess of acid could damage it

NOTE: Filter A for use in G hoods and Filter AD for use in GS hoods

  • Type BE: ideal for inorganic acids and bases as well as volatile compounds of sulphur such as H 2S, H 2SO 4, SO 3. Can be used with organic vapours as the activated carbon is impregnated with metallic compounds and neutralising salts.

NOTE: Filter BE for use in G hoods and Filter BED for use in GS hoods

  • Type F: ideal for formaldehyde and derivatives, as well as organic compounds. The carbon is impregnated with Cu and it should never be used with organic acids.

NOTE: Filter F for use in G hoods and Filter FD for use in GS hoods

  • Type K: ideal for NH 3 and amines. Also good for organic compounds. Impregnated with metallic complexes salts

NOTE: Filter K for use in G hoods and Filter KD for use in GS hoods

  • Type G: specific for radioactive iodine and organic iodine compounds. Includes a HEPA filter for solids (ref. Filter D)

NOTE: Filter G can only be used is GS hoods

  • Type M: specific for Hg (Mercury). It should be fitted with a HEPA filter for solids (ref. Filter D)

NOTE: Filter M can only be used in the GS hoods

  • Filter D: The designation D indicates a HEPA absolute filter class H14 (EN-1822 standard) is fitted as a prefilter below the activated carbon filter required for applications that generate solids and/or aerosols

Modular units: All CRUMA fume hoods can be joined to form multiple units

Movilair: The MOVILAIR portable trolley can be used to easily re-locate any of the CRUMA fume hoods and consist of:

  • Structure made from 2mm. steel plate with acid resistant polymerised Epoxy coating
  • A tray inside the trolley divides the useful space in two, forming a practical shelf and allowing work while seated
  • Transparent polymethyl methacrylate 8mm. walls, highly resistant to acids, fire and knocks
  • Transport trolley with 4 wheels (200 Kg. max.). Two of them with a locking device to ensure stability at all times

 

CRUMA Ductless Powder Weighing Stations

Because of their special handling characteristics, all CRUMA negative pressure powder weighing stations are fitted with a high efficiency HEPA filter. Apart from accurate ventilation and optimum visibility, this powder weighing station is specially designed to guarantee the comfort and safety of the user while manipulating products in its interior.

The powder weighing station version is applicable to the whole range of CRUMA fume hoods with filters offering the powder weighing station that best suits the customer specific needs. All models can also be fitted to the MOVILAIR support trolley for portability. The main specification difference is in the 'air renewals per minute e.g. the CRUMA 870P has 8.4 air renewals per minute. The powder weighing fume hoods versions have a 'P' suffix

 

CRUMA Ductless Fume Hood for Education

CRUMA offers a specially designed fume hood which includes al the safety and protection systems required for professional applications. In addition, this fume hood is fitted with a 15mm. thick transparent polymethyl methylacrylate rear wall that allows teachers and other students to see how the work is carried out inside the hood from any angle. Request for the optional transparent rear wall.


 
Search
Search Site:
Make an Enquiry








Brochures
Progen Focus 2008 Download Adobe Acrobat Reader
Food, Feed and Environmental Analysis Equipment from Progen Scientific Progen Five Star Capital Equipment
To view the brochures you will need to have Acrobat Reader
 
Site Build and Hosting by Wizzy Design Ltd.  |   Design By Website Optimisation  |   Powered By Apex CMS ©
Cruma Ductless Fume Hoods © Progen Scientific Ltd. 2008. All Rights Reserved.